A Dynamic Method for the Calibration of Dye Dilution Curves in a Physiological System

Journal/Book: Brit. Heart J. 28 1966 143.. 1966;

Abstract: From the Institute of Cardiology National Heart Hospital and the Institute of Clinical Research Middlesex Hospital London W. 1 SUMMARY In 41 experiments involving 11 dogs the cardiac output was measured simultaneously by multiple arterial samples (Hamilton et al. 1928) and from a continuously recorded dilution curve using the dynamic method of calibration (Sparling et al. 1960). There was a 95 per cent probability that the results were within -11·0 to +13·5 per cent of each other thus establishing the accuracy of this method of calibration in a physiological system. . . .