Heilpflanzen-Welt - Die Welt der Heilpflanzen!
Heilpflanzen-Welt - Natürlich natürlich!
December 2024

Man Ther. 2001 Nov; 6(4): 229-34.

Development of an instrumented couch to measure forces during manual physiotherapy treatment.

Chiradejnant A, Maher CG, Latimer J.

School of Physiotherapy, Faculty of Health Sciences, The University of Sydney, NSW, Australia. [email protected]

This paper reports on an instrumented couch developed to measure the forces applied during posteroanterior (PA) mobilization to the lumbar spine. The couch was evaluated for reliability by applying known weights to the empty couch, while accuracy was investigated by applying weights both to the empty couch and to a human subject lying on the couch. The equipment appears highly reliable when measuring forces in three directions (ICC(2,1); 99% CI=0.99-1.00) and also demonstrates acceptable accuracy; a maximum error of less than 2% being obtained when measuring forces applied to human subjects. The results demonstrate a highly linear relationship (r2=0.99) between the force readings and the known weights. In conclusion, the instrumented couch demonstrates suitable accuracy and reliability for measuring the force applied to a subject during the application of PA mobilization to the lumbar spine.Copyright 2001 Harcourt Publishers Ltd.